HD-E702-100K40 CLIMATE CHAMBER

CLIMATE CHAMBER (juga dikenal sebagai kamar uji suhu dan kelembaban) menyediakan rentang kontrol suhu dan kelembaban yang luas. Alat ini digunakan untuk mensimulasikan kondisi lingkungan ekstrem seperti suhu tinggi, suhu rendah, atau kelembaban tinggi.

Pelanggan dapat memilih model yang sesuai berdasarkan performa dan kapasitas area pengujian. Kamar uji ini banyak digunakan di berbagai industri, seperti instrumen dan meter elektronik, material baru, kelistrikan, kendaraan, logam, produk elektronik, serta material dirgantara. Peralatan ini membantu menguji ketahanan produk terhadap kondisi lingkungan yang mungkin dialami selama transportasi, penyimpanan, dan penggunaan.

Cooling System

No Description
1 The refrigerating system equips France Tecumsch, and uses environmentally friendly refrigerant (R404a and R23). Adopting binary low temperature circuit system design, it is convenient to use different compressors for work in different temperature regions.
2 Temperature and humidity control using P.I.D + S.S.R system coordinated control with the same channel. It has the function of automatic calculation, which can immediately modify the temperature and humidity change conditions to make the temperature and humidity control more accurate and stable.
3 High-precision controllers of temperature and humidity. The relative temperature and humidity performance is accurately set and displayed with a resolution of 0.1 ℃.

Temperature and Humidification System

No Description
1 Temperature system using Nickel-chromium alloy high-speed heating electric wire;
2 High temperature system is completely independent;
3 The humidity system using stainless steel tube built-in nickel-chromium alloy high-speed heating electric wire and high-temperature magnesium oxide powder, which has fast response and long life;
4 The temperature and humidification system is completely independent, and humidity has automatic water supplement device.

Control System

Component Description
Temp. & Humid. Sensor High precision PT100 Platinum resistance
Controller UE
Display Function a. Temp. & humid. Setting (SV) Practical (PV) value can be displayed directly b. Display numbers, paragraphs, remaining time and cycles, running time c. Program editing and graphic curve display, d. Fixed or program operation status display,
Interface a. RS-232 or R485 port design b. Connected to computer c. Monitor and remote control by computer and Ethernet optional

Protection System

No Description
1 Protection for electrical leakage
2 Over-temperature
3 Compressor overvoltage and overload
4 Heater short circuit
5 Water shortage

Capacity and Material

Category Details
Internal Dimension (W*D*H) 400*500*500mm (100L)
External Dimension (W*D*H) 620*1410*1390mm
Internal Material SUS#304 stainless steel, mirror finished
External Material Cold-rolled steel plate baking

Temperature

Parameter Details
Temperature Range -40℃ ~ +150℃
Temperature Fluctuation ±0.5℃
Temperature Uniformity ≤2℃
Temperature Rate 1℃/min in average (without loading)
3℃/min in average (without loading)
Temperature. &humidity

Humidity

Parameter Details
Humidity Range 20%~98%RH
Humidity Fluctuation ±3%RH
Humidity Uniformity ±2%RH

Power Supply

Power Source: single-phase, AC220V/50Hz

Weight: Approx. 200kg