Deskripsi
Highlighted Features
- Wirelessly transmits real-time data
- Compact size for easy deployment wherever data is needed
- Accepts a wide range of external energy and environmental sensors
- Notifies you of alarm conditions via email or text messages
- Onboard buffer memory helps prevent data loss
- Powerful HOBOware® Pro software for organizing and viewing data (included with HOBO data receiver)
A complete system requires at least one wireless HOBO data node, HOBO data receiver (ZW-RCVR), and HOBOware Pro software (included with ZW-RCVR). Depending on your system deployment, a HOBO data router (ZW-ROUTER) may be required.
In what environment does this data logger operate?
This data logger operates in an indoor environment.
What measurements does this data logger support?
The ZW-005 data logger supports the following measurements: 4-20mA, AC Current, AC Voltage, Air Velocity, Amp Hour (Ah), Amps (A), Carbon Dioxide, Compressed Air Flow, DC Current, DC Voltage, Differential Pressure, Gauge Pressure, Kilowatt Hours (kWh), Kilowatts (kW), Power Factor (PF), Pulse Input, Relative Humidity, Temperature, Volatile Organic Comp., Volt-Amp Reactive, Volt-Amp Reactive hour, Volt-Amps (VA), Volts (V), Water Flow, Watt Hours (Wh) and Watts (W)
The country of origin for this product is the United States. To see the full specifications for this product, please see our product manual found under the documentation tab if available
Buffer memory: 3k measurements
Sample rate: 1 min to 18 hrs
Transmission rate: 2 min and greater
Power options: AC power adapter, Battery Backup
Measurement Range
Temp: -40° to 70° C (-40° to 158° F)
RH: 0 to 100% RH
Analog channels: 0 to 2.5 Vdc; 0 to 5 Vdc (w/CABLE-ADAP5); 0 to 10 Vdc (w/ CABLE-ADAP10)
Pulse channel: (w/CABLE-2.5-STEREO)0 to 65,535 pulses per logging interval
Accuracy
Temperature: ±0.21°C from 0 to 50°C (±0.38°F from 32° to 122°F); see Plot A
RH: ±2.5% from 10%RH to 90%RH typical to a maximum of ±3.5% including hysteresis at 25˚C (77˚F); below 10%RH and above 90%RH ±5% typical
Analog channel: ± 1.544 mV plus 2% of reading typical
Operating Range
Temp: -20 to 50°C (-4 to 122°F) when battery powered, -20 to 70°C (-4 to 158°F) when line powered.
RH: 5%-95% non-condensing
Resolution
Temp: 0.02° C @ 25° C (0.04° F @ 77° F)
RH: 0.05%Analog channel: 0.6 mV
Pulse channel: 1 pulse
Response Time
Temp: 5 min. in air moving 1m/sec (3.3 ft./sec)
RH: 10 min. in air moving 1 m/sec (3.3ft/sec)
Range: Approx. 100m (300ft.) depending on obstructions or interference
Weight: 138g (4.87 oz) with batteries
Dimensions: 96.5 x 108 x 28mm (3.8 x 4.25x 1.1 in)
Radio power: 1.6 mW (2 dBm)
Wireless data standard: IEEE802.15.4 2.4 GHz band
Ulasan
Belum ada ulasan.